Features:
1. LCD display.
2. Double control modes of bed temperature and skin temperature.
3. Double-layer thermostatic cover, automatic air curtain device.
4. Humidity control function.
5. Air&Oxygen blender.
6. Oxygen supply system.
7. Baby cot is infinitely adjustable.
8. 24 kinds of sound and light alarm, comprehensive protection of baby safety.
Technical Parameters:
General Parameters | Item | Specifications |
Power Supply | AC220V, 50/60Hz(Standard) 1320VA; | |
Size(L*W*H) | 950*645*(1250~1450)mm | |
Maximum Heater Output Power(220VAC) | Air Chamber Heater: 330W Water Tank Heater: 155W | |
Auxiliary Network Power Output | AC220V/50Hz, both maximum permissible currents are 220V 1.5A | |
Heater Indication | 0~100%, 10% increasing amount | |
Temperature Control Mode | Bed temperature mode; The skin temperature mode | |
Bed Temperature Control Temperature Range | 25.0~37.0℃; 37.1~39.0℃(Span mode) | |
Skin Temperature Control Range | 34.0~37.0℃; 37.1~38.0℃(Span mode) | |
Skin Temperature Sensor Display Range | 5.0~65.0℃ | |
Heating Time | ≤50min | |
Temperature Variability | ≤0.5℃ | |
Temperature Uniformity(Mattress in horizontal position) | ≤0.8℃ | |
Temperature Uniformity(Mattress in tilted position) | ≤1.0℃ | |
Difference Between Average Incubator Temperature and Control Temperature | ≤1.5℃ | |
Skin Temperature Sensor Accuracy | ±0.3℃ | |
Humidity Control | The running time of the tank after it is filled with distilled water | Humidity below 70%RH, at least 24h |
Water Tank Capacity | 1500ml | |
Humidity Display Range | 0%RH~99%RH | |
Humidity Control Range | CAH%RH~90%RH(1% per adjustment) | |
Humidity Control Accuracy | ±10% | |
Oxygen Concentration Control | Oxygen Concentration Display Range | 0%~99% |
Oxygen Concentration Display Resolution | 1% | |
Oxygen Concentration Control Accuracy | ±5% oxygen volume concentration | |
Oxygen Concentration Setting Range | 21%~60%(1% per adjustment) | |
Oxygen Sensor Service Life | The maximum use time is 10000h at 100% oxygen concentration | |
Transport Storage | Temperature | -20℃~+55℃ |
Atmospheric Pressure Range | 500hPa~1060hPa | |
Relative Humidity | ≤93%RH | |
Packing Size/Gross Weight | 1150*800*1600mm 190kg | |
The Work Environment | Relative Humidity | 30%RH~75%RH |
The Environment Temperature | +20℃~+30℃ | |
Pressure Range | 700hPa~1060hPa | |
Service Altitude | ≤2000m | |
Over Voltage Category | II | |
Pollution Levels | 2 | |
Ambient Air Velocity | ≤0.3m/s | |
Other Indicators | Environmental Noise In Incubator Baby Compartment | No more than 55dB(A) at stable temperature |
Carbon Dioxide Concentration In The Infant Compartment | A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0.5% after stabilization | |
The velocity of air flow over a baby's mattress | ≤0.35m/s | |
General Alarm | Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm | |
The System Alarm | Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure |
Configuration:
Model | Type I+ | Type I | Type II | Type III | Type IV |
Temp. Control | Bed/Skin Temperature | Bed/Skin Temperature | Bed/Skin Temperature | Bed/Skin Temperature | Bed/Skin Temperature |
Display | LCD | LCD | LCD | LCD | LCD |
Bed Temp. Control Range | 25℃~39℃ | 25℃~39℃ | 25℃~39℃ | 25℃~39℃ | 25℃~39℃ |
Skin Temp. Control Range | 34℃~38℃ | 34℃~38℃ | 34℃~38℃ | 34℃~38℃ | / |
Humidity Control Range | CAH%~99%RH | CAH%~99%RH | CAH%~99%RH | / | / |
Oxygen Control Range | 21%~99% | 21%~99% | / | / | / |
Cabinet | Yes | Yes | Yes | Yes | Yes |
Height Adjustable | Yes | / | / | / | / |
Download: